Features:
Model
|
BKE Series
|
Sample
|
Serum, plasma, whole blood, cerebrospinal fluid and dilute urine
|
Measuring Speed
|
≤25s
|
Analysis Method
|
Ion selective electrode (ISE)
|
Sample Volume
|
60~150ul
|
Sample Position
|
30 sample positions, 5 emergency positions, 2 positions
|
Injection mode
|
Manual or automatic injection can be selected
|
Storage
|
Up to 50,000 test results
|
Printer
|
Build-in thermal printer
|
Interface
|
RS232 port for LIS
|
Display
|
7-inch high-definition touch screen
|
Language
|
Chinese and English, other languages can be customized
|
Calibration
|
Automatic and Manual calibration
|
Other Function
|
Sample volume detection
|
Power Supply
|
AC110~240V,50/60Hz, 100W
|
Temperature
|
10~30℃
|
Relative Humidity
|
≤ 80 %
|
Atmospheric Pressure
|
86~106 kPa
|
Items
|
Measuring Range
|
Resolution
|
K+
|
0.5-20.0mmol/L
|
0.01 mmol/L
|
Na+
|
15-200mmol/L
|
0.1 mmol/L
|
Cl-
|
15-200mmol/L
|
0.1 mmol/L
|
Ca2+
|
0.1-6.0mmol/L
|
0.01 mmol/L
|
pH
|
4-9pH
|
0.01 pH
|
Model
|
BKE-A
|
BKE-B
|
BKE-C
|
BKE-D
|
Test Items
|
K+, Na+, Cl-
|
K+, Na+, Cl-
|
K+, Na+, Cl-, iCa2+, nCa2+, TCa2+, pH
|
K+, Na+, Cl-, iCa2+, nCa2+, TCa2+, pH
|
Sample Tray
|
No
|
Yes
|
No
|
Yes
|
External Size(W*D*H)
|
390*200*434mm
|
660*276*434mm
|
390*200*434mm
|
660*276*434mm
|
Net Weight
|
8.2kg
|
11.9kg
|
8.2kg
|
11.9kg
|
Package Size(W*D*H)
|
500×390×530 mm
|
500×390×700 mm
|
500×390×530 mm
|
500×390×700mm
|
Gross Weight
|
13.8kg
|
18.4kg
|
13.8kg
|
18.4kg
|