OLABO 2000 Auto Auto ELISA Processor
Features
1. 1 robotic arm, 2 pipetting probe(10~1000μl).
2. 4 unit 96 well microplates(independent incubating).Modularized automatic control reading and washing system.
Technical Parameters
Model
OLABO2000
Sample Unit
Sample Racks
552 sample positions (tubes 13mm)
Pipetting
8~1000μl
Dispensing System
2 aspirating and dispensing probe (X-Y-Z movement)
Reagent Unit
Reagent Position
23 reagent racks, editable
Pipetting Range
8~1000μl, 1μl stepping
Pipetting Precision
(100μl) CV≤0.5%
Washing Unit
Dispensing Time
4 minutes to whole 96 well microplates
Washing Probes
Two line 8-nozzle manifolds
Wash Containers
Cleaning fluid (15L), waste water (25L), Buffer 1 (5L), Buffer 2 (2L), Buffer 3 (2L), Buffer 4 (2L), with liquid level-sensing, liquid shortage and full alarm
Waste Containers
25L with waste full sensor
Washing Residual
< 2μl
Reading Unit
Reading Channels
8 independent photometric channels, momo and dichromatic reading
Absorbance
0.0~3.0 OD
Spectral Range
400~700mm
Optical Filters
2 standard filters (450, 630mm); 2 more on request (405nm/450nm/492nm/630nm)
Reading Accuracy
±1% from 0~1.5OD, ±2% from 1.5~3.0 OD
Incubation Unit
Incubators
4 independent units
Temperature Range
RT to 45℃
Temperature Accuracy
±0.5℃
Data Management
Software System
Windows 7
LIS System
Bi-direction, support HL7 protocol
Results
Absorbance and results reviewable by software
Printer
External optional
Working Conditions
Power Supply
AC220V±10% 50/60Hz; 110V±10% 60 Hz, 400W
Temperature
10~30℃
Humidity
30~80%
Size & Weight
External Size(W*D*H)
1235*710*880mm
Net Weigth
146kg
Package Size(W*D*H)
1380*860*1435mm
Gross Weight
220kg