OLABO Micropipette
Features
1. Lightweight, ergonomic, low force design
2. Digital display clearly reads volume setting
3. The pipettes cover volume range of 0.1μl to 10ml
4. Easy to calibrate and maintain with tool supplied
5. Design helps avoid repetitive strain injuries
6. Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
7. The low part is available for autoclaving
Technical Parameters
Volume Range
Increment
Test Volume(μl)
Accuracy error
Precision error
%
μl
%
μl
0.1-2.5μl
0.05μl
2.5
2.50%
0.0625
2.00%
0.05
1.25
3.00%
0.0375
3.00%
0.0375
0.25
12.00%
0.03
6.00%
0.015
0.5-10μl
0.1μl
10
1.00%
0.1
0.80%
0.08
5
1.50%
0.075
1.50%
0.075
1
2.50%
0.025
1.50%
0.015
2-20μl
0.5μl
20
0.90%
0.18
0.40%
0.08
10
1.20%
0.12
1.00%
0.1
2
3.00%
0.06
2.00%
0.04
5-50μl
0.5μl
50
0.60%
0.3
0.30%
0.15
25
0.90%
0.225
0.60%
0.15
5
2.00%
0.1
2.00%
0.1
10-100μl
1μl
100
0.80%
0.8
0.15%
0.15
50
1.00%
0.5
0.40%
0.2
10
3.00%
0.3
1.50%
0.15
20-200μl
1μl
200
0.60%
1.2
0.15%
0.3
100
0.80%
0.8
0.30%
0.3
20
3.00%
0.6
1.00%
0.2
50-200μl
1μl
200
0.60%
1.2
0.15%
0.3
100
0.80%
0.8
0.30%
0.3
50
1.00%
0.5
0.40%
0.2
100-1000μl
5μl
1000
0.60%
6
0.20%
2
500
0.70%
3.5
0.25%
1.25
100
2.00%
2
0.70%
0.7
200-1000μl
5μl
1000
0.60%
6
0.20%
2
500
0.70%
3.5
0.25%
1.25
200
0.90%
1.8
0.30%
0.6
1000-5000μl
50μl
5000
0.50%
25
0.15%
7.5
2500
0.60%
15
0.30%
7.5
1000
0.70%
7
0.30%
3
2-10ml
0.1ml
10ml
0.60%
60
0.20%
20
5ml
1.20%
60
0.30%
15
2ml
3.00%
60
0.60%
12